抖圈为du而生

N1008A(800Vp-p,50MHz) differential probe 1% accuracy

WeChat
Sina Weibo
QQ
QQ space
Douban
Baidu Post Bar
Bandwidth 50MHz Voltage overload protection Input impedance 2MΩ//2.5pF each side ground 1% accuracy,compatible with any oscilloscope,standard BNC interface. 2 attenuation ratio range,easy to operating.
Brand PINTECH
Model N1008A
DetailDimension drawingFrequency diagramAccessory detail

Model

N1008A

Bandwidth (-3dB)

DC-50MHz

Attenuation Ratio

100X,,,,,,10X

Accuracy

±1%

Input Impedance

Between inputs:4MΩ//1.25pF

Each input to ground:2MΩ//2.5pF  

Rise time

7ns

-Output(into 50kΩ load)

<7V

-Offset(typical)

<±5mV

oscilloscope input impedance

1MΩ

CMRR(Typical)

60Hz:>80dB;;;;;;;100kHz:>50dB

Test voltage Vp-p

10X:80V

100X:800V

Input DC voltage

10X:40V

100X:400V

Differential mode voltage Vp

10X:40V

100X:400V

Common mode voltage(RMS)

280V

RMS voltage(RMS)

10X:28V

100X:280V

Noise(RMS)

10X:<10mV

100X:<50mV

Length of BNC Cable

90cm

Length of Input Leads

30cm

Working Temperature

0~50℃

Working   Humidity

20~75%RH

Weight

284g

Dimension (LxWxH)

166.5*70.7*30mm

Manual measurement, error ±2mm

幼白机表形尺寸图.jpg


N英.jpg

Jasmine
Tel:+86 13825053608
Jasmine
Jasmine
WeChat
Skype
sales-Pintech
【网站地图】【sitemap】